Dr. SPECKLE Model:SM01VS08(Old Product)

‹Type2› Universal measurement: SM01VS08

  • Measurement head tailored to users’ request
  • Interface box connected to control PC
  • We can offer components such as test bench, stages, etc.

[Samples] projectors (rear or front), lasers, light sources, screen illuminations, HUD, etc.

Measurement head

SM01VS08-img02
Interface box

Specifications

■Measurement

Speckle contrast C=σ/Ibar
Dynamic range 0.01~1.00

■Measurement Conditions (Measurement Object)

Luminous Flux 10lm (lumens)or over (monochromatic)
Irradiation screenφ ~φ25mm in diamater or over
Wave length Visible region
Projection image Light beam with uniform light density distribution

 ■Operating Ambient

Operating Temp. 15~35 Centigrade
Operating Humidity 10~70%(No dew condensation)
Storage Temp. 0~40 Centigrade
Storage Humidity 10~70%(No dew condensation)

■Aperture Size of CCD Camera

Aperture diameter 0.8mm, 1.0mm ,1.2mm ,1.5mm (selectable)

■Electrical Specifications

Supply Voltage ACSingle Phase 85~264V
Frequency 47~63Hz
Consumption 125W

■Requirement for control PC

Type DOS/V compatible PC
 OS Microsoft® Windows® XP、Vista、7
 CPU  Intel® Core™2 Duo T8100(2.10GHz)or higher recommended
 Memory 1GB or over (2GB or over recommended) 
 Display  1280×800 or higher (1,600 x 900 recommended)
 Interface  USB2.0 at least 3ports
Storage drive  CD-ROM Drive (needed for Measurement Software installation)

◊ Customizations are available in each subjects